Defibrillator Monitor HYZ-9000C&HYZ-8000C
Physical characteristics:
Main unit: Portable
Dimensions:L320xW205xH410
Weight:9.8kg
Technical specifications |
|
Monophasic
|
Biphasic
|
HYZ-9000C Defi-monitor |
HYZ-8000C Defi-monitor |
DEFIBRILLATOR. |
|
(1) Defibrillator type |
Manual, Synchronized, Asynchronized, |
(2) Energy accuracy |
<±1% |
(3) Energy selection |
0,3,5,7,10,20,30,50,100,200,300,360,Joules(nominal at a resistance of 50) |
(4) Charging time |
Maximum 10 seconds to 360 Joules |
(5) Standard Adult/Pediatric Paddles |
Reusable external adult paddles |
Power requirements. |
|
(1) AC Power supply |
100V~240V,50Hz/60 Hz |
(2) Vehicle Voltage(Option) |
DC 12V |
Internal Battery Backup |
|
(1)Battery type |
(1) Rechargeable lead-acid battery (12V 2.0Ah X2) |
(2) Charging time |
Time-Minimum of 4 hours for full charging |
(3) Battery charging |
During AC Power Supply operation of the device. |
(4) Charging /Low Battery Indication |
Red LED |
(5) Battery Capacity |
100% of the battery will provide 120-min monitoring; and 35 discharges at 360 joules. |
Display |
|
LCD Dimensions |
7-inch |
Type |
High resolution true-color graphic |
Resolution |
480X234 |
ECG |
|
ECG lead: |
Ⅰ, Ⅱ, Ⅲ, AVR, AVL, AVF, V1~V6 |
Protection: |
With stand 4000vAc/50Hz voltage in isolation and work against electrosurgical knife and defibrillation |
Respiration |
|
Measurement method: |
Thoracic impedance |
NIBP |
|
Working mode: |
Manual/Automatic/stat(5 minutes) |
Method |
Automatic oscillation |
Cuff type |
standard Adult; optional Pediatric/Neonatal |
Temperature |
|
Measurement range |
20~45℃ |
Resolution |
0.1℃ |
Pulse Oximetry(spO2) |
|
Display |
spo2 value, pulse histogram, waveform, Pulse |
Range |
0~99%for adult, pediatric and neonatal patients |
Probe |
standard adult finger clip |
optional pediatric Y-type clip and neonatal wrap |
|
Pulse range |
|
Pulse range |
0-255bpm |
Standard parameter & configuration |
|
ECG/ RESP/SPO2/TEMP/PR/NIBP/DEFIBRILLATOR |
|
Optional |
|
Recorder, Li-ion battery (11.1V 4Ah X2) |