HYZ-SS90/160 Light Double Arm Revolving Pendant

HYZ-SS90/160

Recommended Products
    Product information

     Installation Instructions:

    The related contractor should be responsible for the gas piping/power line/communication line:

    1 .Laying medical gas supplying line to the anchor point of the pendant, putting the gas piping into the pendant, angle branch radius:R>=30mm.The diameter requirements for the gas piping: 10mm(suction), others 8mm(oxygen etc).

    2. For laying power lines,communication lines to the positioning point of the pendant,the reserved length shall be guaranteed <1 m after spreading to the ground.For power line: 3x2.5mm2, 1x6mm2

     According to customer's need or customize according to locate