HYZ-SD90/160 Light Single Arm Revolving Pendant

HYZ-SD90/160

Recommended Products
    Product information

    Installation Instructions:

    The related contractor should be responsible for the gas piping/power line/communication line:

    1. Laying medical gas supplying line to the anchor point of the pendant, putting the gas piping into the pendant, angle branch radius: R>=30mm. The diameter requirements for the gas piping: 10mm(suction), others 8mm(oxygen etc).

    2. For laying power lines, communication lines to the positioning point of the pendant, the reserved length shall be guaranteed <1 m after spreading to the ground. For power line: 3x2.5mm2, 1x6mm2 

    According to customer's need or customize according to locate