Infusion Pump HYZ-820/HYZ-820D With drug library
Infusion Pump HYZ-820
HYZ-820D With drug library
1. Bubble detector and pressure sensor can precisely detect tube air bubbles
2. Automatically calibration and compatible with any brand infusion sets.
3. Unique human voice alarm system and accurate injection rate make the injection process safer and more reliable
4. HD LCD display working status.
5. Working mode: ml/h and drop/min, and these two modes can switch freely.
6. KVO and Bolus functions.
Technical parameters of HYZ-820/820D infusion pump:
Rate: |
1ml/h~1200ml/h |
Special function |
HYZ-820D With drug library |
Flow Rate Accuracy: |
Within ±5% (after correct calibration) |
Mechanical precision: |
Within ±2% |
Purge Rate: |
100ml/h~1,000ml/h (100 ml/h step) |
Volume Limit: |
1ml~9999ml |
Total Infusion Volume: |
0ml~9999.9ml |
KVO rate: |
1ml/h~5 ml/h (1ml/h step) |
Occlusion: |
High: 800mmHg ±200mmHg (106.7kPa±26.7kPa Medium: 500mmHg ±100mmHg(66.7kPa±13.3kPa Low: 300mmHg ±100mmHg (40.7kPa±13.3kPa ) |
Unique audible and visible alarm |
Human voice alarm for End of injection, occlusion, door open, bubbles in the tube, wrong setting, low battery, AC power pulled out etc. |
Power Source: |
AC 100V~240V, 50/60Hz; Internal rechargeable Li battery, capacity≥1,600mAh, 4 hours internal battery backup |
Bubble detector: |
Ultrasonic wave detector; detection sensitivity ≥25μL |
Power Consumption: |
30VA |
Operating Condition: |
Ambient temperature: +5℃~+40℃; Relative humidity: 20~90% Atmospheric pressure: 86.0kpa~106.0kpa |
Transport & Storage Condition: |
Ambient temperature: -30℃~+55℃ Relative humidity: ≤95% |
Equipment Classification: |
Class II, internal power supply, Type CF |
IP Classification: |
IPX4 |
Dimension: |
140mm (L) × 157mm (W) × 220mm (H) |
Weight: |
1.8kg (net weight) |