img

HYZ-810/ HYZ-810D syringe pump

.HYZ-810

HYZ-810D With drug library and 2000 records.
2.Freely Stackable: Users can freely stack one syringe pump onto another to provide multiple Solutions, which have a wide range of clinical applications.
3.Transfer the message between the Nurse-call System and the pumps.
4.Three working modes: rate mode, time-volume mode, dosage-weight mode.
5.Unique Human Voice Alarm system and accurate injection rate make the injection process safer and more reliable.
6.Have store many brands syringes ,Automatically calibration and compatible with any brand syringes.
7.Large LCD screen displays working status.
8.KVO and Bolus functions.

Recommended Products
    Product information

    Technical parameters of HYZ-810 / 810D syringe pump:

    Unique Human Voice Alarm system

    The device triggered the human voice prompt and stop injection automatically when there is any malfunction, making the infusion process safer and more reliable.

    Special function

    HYZ-810D With drug library

    Rate:

    50/60ml Syringe:  0.1ml/h999.9ml/h   (0.1ml/h step)
    1,000ml/h1,500ml/h   (1ml/h  step)
    30ml Syringe: 0.1ml/h900.0ml/h    (0.1ml/h step)
    20ml Syringe:  0.1ml/h600.0ml/h   (0.1ml/h step)
    10ml Syringe:  0.1ml/h300.0ml/h   (0.1ml/h step)

    Flow Rate Accuracy:

    Within ±3% (after correct calibration)

    Mechanical precision:

    Within ±2%

    Bolus Rate:

    50/60ml Syringe:  1,200ml/h
    30 ml Syringe:  720ml/h
    20 ml Syringe:  480ml/h
    10 ml Syringe:  240ml/h

    Purge Rate:

    50/60ml Syringe:  1,500ml/h
    30ml Syringe:  900ml/h
    20ml Syringe:  600ml/h
    10ml Syringe:  300ml/h

    Volume Limit:

    0.1ml9999.9ml

    Total Injection
    Volume:

    0.1ml9999.9ml (0.1ml step)

    Occlusion:

    High:   800mmHg ±200mmHg     (106.7kPa±26.7kPa
    Medium: 50/600mmHg ±100mmHg  (66.7kPa±13.3kPa
    Low:    300mmHg ±100mmHg    (40.7kPa±13.3kPa )

    Alarms:

    Injection soon finish, end of injection,occlusion, improperly installation of syringe, wrong setting, low battery, syringe loose etc.

    Power Source:

    AC 100V240V, 50/60Hz; Internal rechargeable Li battery, capacity≥1,600mAh, 4 hours internal battery backup

    KVO rate:

    1ml/h

    Fuse:

    F1AL/250/60V, 2pcs inside

    Power Consumption:

    30VA

    IP Classification:

    IPX4

    Equipment Classification:

    Class I, internal power supply, Type CF

    Operating Condition:

    Ambient temperature: +5℃+40℃;

    Relative humidity:

    2090%

    Transport & Storage Condition:

    Ambient temperature: -30℃+55℃

    Relative humidity:

    ≤95%

    Dimension:

    280mm (L) × 210mm (W) × 130mm (H)

    Weight:

    2.2kg (net weight)

    © HYZ Medical Equipment Limited. www.hyzmed.com 翰智医疗 All rights reserved